Surgeon Estimations of Acetabular Cup Orientation Using Intraoperative Fluoroscopic Imagining Are Unreliable.
Document Type
Article
Publication Date
4-2023
Abstract
BACKGROUND: Accurate acetabular cup orientation is associated with decreased revision rates and improved outcomes of primary total hip arthroplasty. This study assesses surgeon's ability to estimate both the acetabular component inclination and anteversion angles via intraoperative fluoroscopy (IF) images.
METHODS: We surveyed orthopedic surgeons to estimate acetabular component inclination and anteversion based on 20 IF images of total hip arthroplasty through a direct anterior approach. Postoperative computed-tomography scans were used to calculate the true inclination and anteversion component angles. The absolute difference between the true and estimated values was calculated to determine the mean and standard deviation of the survey results. Interrater reliability was determined through interclass correlation coefficients.
RESULTS: A majority of surgeons preferred the direct anterior approach (83.3%) and utilized IF during surgery (70%). Surgeons surveyed were on average 5.9° away from the true value of inclination (standard deviation = 4.7) and 8.8° away from the true value of anteversion (standard deviation = 6.0). Respondents were within 5° of both inclination and anteversion in 19.7% of cases, and within 10° in 57.3% of cases. All surgeons were determined to have poor reliability in estimating anteversion (interclass correlation coefficient < 0.5). Only 2 surgeons were determined to have moderate reliability when estimating inclination.
CONCLUSIONS: Surgeons, when solely relying on IF for the estimation of anteversion and inclination, are unreliable. Utilization of other techniques in conjunction with IF would improve observer reliability.
Publication Title
Arthroplast Today
Volume
20
First Page
101109
Last Page
101109
Recommended Citation
Brush, P., Santana, A., Toci, G., Slotkin, E., Solomon, M., Jones, T., & Saxena, A. (2023). Surgeon Estimations of Acetabular Cup Orientation Using Intraoperative Fluoroscopic Imagining Are Unreliable.. Arthroplast Today, 20, 101109-101109. https://doi.org/https://doi.org/10.1016/j.artd.2023.101109